How does Basal-IQ technology work?
Basal-IQ technology uses a simple linear regression algorithm that uses Dexcom G6 CGM values to predict glucose levels 30 minutes ahead based on 3 of the last 4 consecutive CGM readings. If the glucose level is predicted to be less than 80 mg/dL, or if a CGM reading falls below 70 mg/dL, insulin delivery is suspended. Insulin delivery resumes as soon as sensor glucose values begin to rise. Insulin may be suspended for a minimum of 5 minutes and a maximum of 2 hours within a 2.5-hour rolling window.